Rapidox 1100 Argon Analyser

Product Information

The Rapidox 1100-G argon (Ar) gas analyser is the company’s most cost-effective and versatile argon gas analyser fitted with a Thermal Conductivity gas Detector.

The Thermal Conductivity Detector (TCD) is a state-of-the art German engineered sensor and is the most common and preferred solution for argon measurement providing fast and accurate gas analysis over the range 0% to 100% Ar in simple binary gas systems, where the argon is mixed with another background gas or air/nitrogen. It is particularly suitable for aggressive industrial applications as long as the sample gas is filtered for dust and particles and is non-condensing.  The sensor has a 50°C heated sensor chamber to overcome measurement drift caused by ambient temperature changes.  A common application is measuring the composition of Ar/H2 welding gas mixtures.

Read More
Categories: ,

Product Information

TCD sensors (also known as Katharometers) are highly suitable for measuring gases that have a thermal conductivity value significantly different to air, as long as the gas mixture is simple.  The sensor can be factory calibrated for a specific gas mixture such as argon in hydrogen for example.

Please note that TCD sensors are not very reliable in complex gas mixtures where there are several gases with different thermal properties.  The TCD measurement principle is an indirect  measurement of the overall thermal conductivity of a gas and this is interpreted as a composition based on a factory calibration. The standard R1100-J is available in various ranges 0-5%, 0-10% 0-50% 0-80% and 0-100% Ar selected at the time of ordering.  The standard range however is 0-100% in a background of air/nitrogen, with a choice of gas fittings on the front or rear of the instrument.  Upgrades to Swagelok fittings (1/4” or 6mm) are available at the time of order.  Each version is available with a choice of internal electric sample pump or vacuum ejector pump.

Special variations of the R1100-J model include:

1)            A stainless steel IP54 instrument version with 316 stainless steel instrument case and sealed gasket joints, designed primarily for the food and pharma industries.

2)            A 4U bezel mounted version compatible with 19” rack mount systems.

Finally, the Rapidox 1100 argon analyser with EMC Directive 2004 / 108 / EC. UL/ETL Certification Number: UL-61010-1.

Features

  • German engineered TCD sensor with specific Ar ranges available meaning more flexibility for the user
  • Thermally stable in all atmospheric conditions as the sensor is heated to a constant 50°C
  • Thanks to internal pressure & temperature correction technology, you can be assured of fast and accurate results every time
  • Simple span-gas calibration procedure (with appropriate calibration gas) for after-sale longevity
  • Low maintenance and long-life sensor expectancy – standards expected from a high quality analyser
  • Two fully programmable alarms with open or closed relay for a complete monitoring solution
  • Fully programmable digital and analogue outputs to further enhance the Rapidox 1100’s versatility
  • Software package includes full control of the analyser, live graphing & Excel compatible data-logging
  • Operates on any worldwide mains voltage (90-260VAC) resulting in ease of use wherever you go
  • Options of front or rear gas fittings and bezel mounts (specify on order)
  • The choice between a long-life variable speed pump or vacuum ejector option
  • For more peace of mind, a password protection feature comes as standard
  • Complies with EMC Directive 2004 / 108 / EC. UL/ETL Certification Number: UL-61010-1

Product Specification

Type Measurement
Ar Sensor Range0-10% or 0-100%
Ar Sensor Accuracy±1% full scale
Ar Sensor ResponseApproximately 20 sec for a 90% response
Sample Connections4mm ID / 6mm OD nipple type Rectus or Swagelock Front or rear positioning
CalibrationRequires 2 user selectable gas compositions Normally zero and span
Warm-up Time5 minutes
Voltage Outputs0-10V, user programmable
Current Outputs4-20mA linear, user programmable
Digital OutputsRS232 (RS485 option available): data streamed on demand/Modbus RTU/Ethernet
Max Gas Pressure700-1300 mbar absolute
Max Gas Sample Temperature50°C
Operating Temperature5°C to 35°C
Pump OptionLong life diaphragm pump variable speed 0-1.2 litres per min
Ejector OptionVacuum ejector fitted, running off inlet pressure
Display20 x 4 character OLED display
Supply Voltage90-260VAC, 50/60Hz
Analyser DimensionsBench: 150mm(H) x 247mm(W) x 250mm(D) Panel: 300 x 4U (177mm(H) x 300mm(W)) Multiplex: 150mm(H) x 263mm(W) x 250mm(D)
Weight3.5kg (4kg with bezel)

Customer Testimonials

“We have always been impressed with both the quality of the products and the professionalism shown from the staff at Cambridge Sensotec. We have used their gas analysis technology in our welding gloveboxes for many years to give our products the cutting edge. We would highly recommend Cambridge Sensotec not only for their product quality, but also their after sales service.”

Ricky Hart, Weldlogic Europe Ltd.