Rapidox 1100 Helium Analyser
Product Information
The Rapidox 1100-Q helium (He) gas analyser is the company’s most cost-effective and versatile helium gas analyser fitted with a Thermal Conductivity gas Detector.
The Thermal Conductivity Detector (TCD) is a state-of-the art German engineered sensor and is the most common and preferred solution for helium measurement providing fast and accurate gas analysis over the range 0-10% and 0%-100% He in simple binary gas systems where the helium is mixed with another gas or air. It is particularly suitable for aggressive industrial applications as long as the sample gas is filtered for dust and particles and is non-condensing. The sensor has a 50°C heated sensor chamber to overcome drift caused by ambient temperature changes.
Read MoreProduct Information
TCD sensors (also known as Katharometers) are highly suitable for measuring gases that have a thermal conductivity value significantly different to air, as long as the gas mixture is simple. Examples include He in mixtures of nitrogen. The helium analyser can be factory calibrated for a specific gas mixture such as helium in argon for example.
Please note that TCD sensors are not very reliable in complex gas mixtures where there are several gases with different thermal properties. The TCD is an indirect sensor that measures the overall thermal conductivity of a gas and interprets this as a composition based on a factory calibration. The standard R1100-Q helium analyser is available in various ranges of He selected at the time of ordering. The standard ranges are 0-10% and 0-100% in a background of nitrogen, with a choice of gas fittings on the front or rear of the instrument. Upgrades to Swagelok fittings (1/4” or 6mm) are available at the time of order. Each version is available with a choice of internal electric sample pump or vacuum ejector pump.
Special variations of the R1100-Q helium analyser include:
1) A stainless steel IP54 instrument version with 316 stainless steel instrument case and sealed gasket joints, designed primarily for the welding, food and pharma industries.
2) A 4U bezel mounted version compatible with 19” rack mount systems.
Finally, the Rapidox 1100 helium analyser complies with EMC Directive 2004 / 108 / EC. UL/ETL Certification Number: UL-61010-1.
Features
- German engineered TCD sensor with specific He ranges available meaning more flexibility for the user
- Thermally stable in all atmospheric conditions as the sensor is heated to a constant 50°C
- Thanks to internal pressure & temperature correction technology, you can be assured of fast and accurate results every time
- Simple span-gas calibration procedure (with appropriate calibration gas) for after-sale longevity
- Low maintenance and long-life sensor expectancy – standards expected from a high quality analyser
- Two fully programmable alarms with open or closed relay for a complete monitoring solution
- Fully programmable digital and analogue outputs to further enhance the Rapidox 1100’s versatility
- Software package includes full control of the analyser, live graphing & Excel compatible data-logging
- Operates on any worldwide mains voltage (90-260VAC) resulting in ease of use wherever you go
- Options of front or rear gas fittings and bezel mounts (specify on order)
- The choice between a long-life variable speed pump or vacuum ejector option
- For more peace of mind, a password protection feature comes as standard
- Complies with EMC Directive 2004 / 108 / EC. UL/ETL Certification Number: UL-61010-1
Product Resources
Product Specification
Type | Measurement |
---|---|
He Sensor Range | 0-10% or 0-100% |
He Sensor Accuracy | ± 1% full scale accuracy |
He Sensor Response | Approximately 20 sec for a 90% response |
He Sensor Life Expectancy | >5 years |
Sample Connections | 4mm ID / 6mm OD nipple type Rectus or Swagelock Front or rear positioning |
Calibration | Requires 2 user selectable gas compositions Normally zero and span |
Warm-up Time | 5 minutes |
Voltage Outputs | 0-10V, user programmable |
Current Outputs | 4-20mA linear, user programmable |
Digital Outputs | RS232 (RS485 option available): data streamed on demand/Modbus RTU/Ethernet |
Max Gas Pressure | 500-1500 mbar absolute |
Max Gas Sample Temperature | 50°C |
Operating Temperature | 5°C to 35°C |
Pump Option | Long life diaphragm pump variable speed 0-1.2 litres per min |
Ejector Option | Vacuum ejector fitted, running off inlet pressure |
Display | 20 x 4 character OLED display |
Supply Voltage | 90-260VAC, 50/60Hz |
Analyser Dimensions | Bench: 150mm(H) x 247mm(W) x 250mm(D) Panel: 300 x 4U (177mm(H) x 300mm(W)) Multiplex: 150mm(H) x 263mm(W) x 250mm(D) |
Weight | 3.5kg (4kg with bezel) |