Rapidox 7100 Blue Hydrogen Analyser
Product Information
The Rapidox 7100 Blue Hydrogen (H2) Analyser is a high specification rack
mountable, or bench top instrument designed for the analysis and calculation of calorific
value for hydrogen produced from biomass gasification, zero-CO2-emission methane
pyrolysis, suitable either in a research laboratory or on-site.
Product Information
The Rapidox 7100 Blue Hydrogen Analyser can be used in a laboratory or onsite
to analyse the output of your hydrogen production process. Using a
combination of Thermal Conductivity (TCD), Infra-Red (IR), Tuneable Laser
Diode (TLD) and Electrochemical (EC) sensor technology, the Rapidox
7100 offers an accurate analysis that can be relied upon.
The analyser will measure the gas composition by providing % or ppm values
for Hydrogen (H2), Methane (CH4), Carbon Dioxide (CO2), Carbon
Monoxide (CO) and Oxygen (O2). It also has an optional upgrade to include
a dewpoint sensor to measure the moisture content in the gas sample.
With our bespoke design of the analyser, we can include a range of other
sensors if your production methods have any other gases that
require monitoring. Our customer-designed firmware for the analyser will
also calculate and record the Gross Calorific Value (GCV) of the gas
throughout the measurement.
The Rapidox 7100 Blue Hydrogen Analyser is available as either a
benchtop version for the laboratory or in a 19″ rack mountable case
which can be housed in a suitable enclosure for use in an external or
hazardous area.
The Rapidox 7100 Blue H2 Analyser is highly configurable to suit
individual H2 production requirements.
Features
- Modular design allows for bespoke sensor combinations upon request
- 7” (180mm) full-colour touch screen interface with soft menu keys for a better user experience
- Permanent or sampling modes available for extra functionality
- All data is continuously logged and Excel compatible data can be downloaded via a USB memory stick
- Operates on any worldwide mains voltage (90-260VAC)
- Two fully programmable alarm settings
- Multi-language (Please contact for details)
- Password protection feature for extra security
Product Resources
Product Specification
Type | Measurement |
---|---|
H2 (hydrogen) | 0-100% TCD sensor with background gas correction. ±1% full scale accuracy. 0.1% resolution |
CO (carbon dioxide) | 0-2000ppm, 0-2%, 0-10% or 0-100% Infra-red sensor. ±1% full scale accuracy. 0.1% resolution |
CH4 (methane) | 0-5% (LEL) or 0-100% TLD sensor. ±1% full scale accuracy. 0.1% resolution |
CO2 (carbon monoxide) | 0-5000ppm, 0-5%, 0-20% or 0-100% Infra-red sensor. ±1% full scale accuracy. 0.1% resolution |
O2 (oxygen) | 0-30% or 0-100% electrochemical sensor. ±1% full scale accuracy. 0.01% resolution |
H2O Dewpoint (optional) | -60°C to +20°C or -100°C to +20°C Dewpoint sensor |
Other gases | Contact us for further information |
Normal Operating Conditions | Temperature 0°C to 40°C / Humidity 10 - 90% RH / Pressure 900 to 1100 mbar absolute |
Sampling | Fixed or continuous sampling modes |
Warm-up Time | 3-4 minutes at 20°C |
Voltage Outputs | 0-10V, user programmable |
Operating temperature | -10°C to 40°C |
Current Outputs | 4-20mA linear, user programmable |
Digital Outputs | RS232 (RS485 option available) Data streamed on demand. Modbus RTU/Ethernet |
Data Output | Excel compatible data via USB memory stick |
Alarms | Relay circuits, user programmable |
Sample Connections | 6mm OD or 1/4" Swagelok fittings. Rear positioning |
Calibration | Calibration by the user using zero and span gases An optional Auto-Cal system can be installed on certain models |
Display | 7" (180mm) full-colour LCD with touchscreen operation; resolution 0.01ppm or 0.01% |
Supply Voltage | 90-260 VAC, 50/60Hz |
Analyser Dimensions | Rack Mount: 132mm(H) x 482mm(W) x 365mm(D) / Benchtop: 180mm (H) x 570mm (W) x 345mm (D) |
Weight | Rack Mount: 6.5kg / Benchtop: 6.5kg |